Namibia High-Temperature Thermal Conductivity Tester (Guarded Hot Plate Method)
Product Description
Xiangtan Xiangke DRS Series High-Temperature Thermal Conductivity Tester (Guarded Hot Plate Method)
I. Overview
This instrument is developed based on our factory's DRS-2 water flow plate method thermal conductivity tester. Compared with DRS-2, it improves resolution and automation, shortens testing time, is easier to maintain, and has more reliable performance. This instrument is suitable for testing the thermal conductivity of refractory insulation, ceramic fiber, felt, textiles, plates, and bricks at different temperatures. It is widely used in colleges and universities, research institutions, quality inspection departments, and material analysis and testing in manufacturing plants.
This instrument refers to the following standards: ASTM C518-04 Test Method for Determining Steady-State Thermal Flux and Thermal Transmission Properties by Means of a Heat Flow Meter; GB/T 10295-2008 Determination of Steady-State Thermal Resistance and Related Properties of Thermal Insulation Materials (Heat Flow Meter Method); YB/T 4130-2005 Refractory Materials Thermal Conductivity Test Method (Water Flow Plate Method); GB/T 17911-2018 Refractory Ceramic Fiber Products Thermal Conductivity Test Method, etc.
II. Xiangtan Xiangke DRS Series High-Temperature Thermal Conductivity Tester (Guarded Hot Plate Method) Main Technical Parameters
The DRS series instruments are mainly used for the thermal conductivity detection of thermal insulation materials at high temperatures. DRS-3B has a vacuum atmosphere protection function to prevent sample oxidation.
Instrument Model Specifications | DRS-3 | DRS-3A | DRS-3B |
1. Thermal Conductivity Range | 0.0010-3W/mk | 0.0010-3W/mk | 0.0010-3W/mk |
2. Testing Accuracy | ±5% | ±5% | ±5% |
3. Hot Surface Temperature Range | 100-1200℃ | 100-1200℃ | 100-1200℃ |
4. Calorimetric Temperature Control Method | Constant Temperature Water Bath | Semiconductor Refrigeration | Semiconductor Refrigeration |
5. Heat Protection Temperature Control Method | Semiconductor Refrigeration | Semiconductor Refrigeration | Semiconductor Refrigeration |
6. Calorimetric Plate Size | 75×75 (mm) | 75×75 (mm) | 75×75 (mm) |
7. Heat Flow Meter Size | 75×75 (mm) | 75×75 (mm) | 75×75 (mm) |
8. Number of Sample Temperature Measurement Points | 4 | 4 | 4 |
9. Sample Atmosphere Environment | Air | Air | -0.095MPa; Atmosphere Protection |
10. Sample Size | 200×200~250×250 (mm); Thickness 10~100 (mm) Note: 300×300 (mm) specifications can be customized according to customer requirements. | ||
11. Measurement and Control Method | Fully automatic computer testing and data processing, and can generate test reports and print output. | ||
12. Main Unit Power Supply | Voltage 220V; Frequency 50Hz; Total Power ≤5KW | ||
13. Instrument Configuration | Includes one DRS-3 testing host, one software set, one set of testing thermocouples, one high-precision constant temperature water bath, and one desktop computer (can be provided by the purchaser). | Includes one DRS-3A testing host, one software set, one set of testing thermocouples, one roll of rubber tubing (instrument needs to be connected to tap water), and one desktop computer (can be provided by the purchaser). | Includes one DRS-3B testing host, one software set, one set of testing thermocouples, one roll of rubber tubing (instrument needs to be connected to tap water), one vacuum sealing device, one vacuum pump, and one desktop computer (can be provided by the purchaser). |
14. Optional | Multi-layer testing can be optionally configured according to customer requirements. Each additional layer requires an additional testing thermocouple, and the cost will be calculated separately. | ||
Both DRS-4 and DRS-4A are electrically pressurized and automatically measure thickness. The difference is that the DRS-4 hot plate uses heat-resistant stainless steel, with a slightly lower operating temperature, while the DRS-4A hot plate uses silicon carbide, with slightly lower pressure resistance, and the thickness measurement requires an additional thickness correction system, but the hot surface operating temperature is higher.
Instrument Model Specifications | DRS-4 | DRS-4A |
1. Thermal Conductivity Range | 0.001-3W/mk | 0.001-3W/mk |
2. Testing Accuracy | Better than 5% | Better than 5% |
3. Hot Surface Temperature Range | 200-800℃ | 200-1200℃ |
4. Calorimetric Temperature Control Method | Semiconductor Refrigeration | Semiconductor Refrigeration |
5. Heat Protection Temperature Control Method | Semiconductor Refrigeration | Semiconductor Refrigeration |
6. Calorimetric Plate Size | 75×75 (mm) | 75×75 (mm) |
7. Heat Flow Meter Size | 75×75 (mm) | 75×75 (mm) |
8. Number of Sample Temperature Measurement Points | 4 | 4 |
9. Loading and Unloading Method | Electric | Electric |
10. Pressure Range | 0-1000 (N) | 0-500 (N) |
11. Thickness Measurement Range | 0-120 (mm) | 0-120 (mm) |
12. Thickness Correction | Experimental Correction | Displacement Meter Instantaneous Correction |
13. Sample Atmosphere Environment | Air | Air |
14. Sample Size | 200×200~250×250 (mm); Thickness 10~100 (mm) | |
15. Measurement and Control Method | Fully automatic computer testing and data processing, and can generate test reports and print output. | |
16. Main Unit Power Supply | Voltage 220V; Frequency 50Hz; Total Power ≤5KW | |
17. Instrument Configuration | Includes one DRS-4 testing host, one software set, one set of testing thermocouples, one roll of rubber tubing (instrument needs to be connected to tap water), and one desktop computer (can be provided by the purchaser). | Includes one DRS-4A testing host, one software set, one set of testing thermocouples, one roll of rubber tubing (instrument needs to be connected to tap water), and one desktop computer (can be provided by the purchaser). |
18. Optional Item One | A constant temperature water bath can be selected according to customer requirements, and the circulating water in the constant temperature water bath replaces tap water. | |
19. Optional Item Two | Multi-layer testing can be optionally configured according to customer requirements. Each additional layer requires an additional testing thermocouple, and the cost will be calculated separately. | |
Note: Hot surface temperature specifications of 1500℃ can also be customized according to user requirements.