Namibia High-Temperature Flat Plate Thermal Conductivity Meter, Thermal Conductivity Analyzer

Xiangtan Xiangke Xiangke DRS-III High-Temperature Flat Plate Thermal Conductivity Meter, Thermal Conductivity Analyzer (Guarded Hot Plate Method). This instrument uses the guarded hot plate method to measure the thermal conductivity of insulation materials at high temperatures. It is suitable for testing the thermal conductivity of refractory insulation, ceramic fiber, glass fiber, felt, textiles, plates, and bricks at different temperatures.

Product Description

Xiangke DRS-III High-temperature flat plate thermal conductivity meter, thermal conductivity analyzer (Guarded heat flow meter method), thermal conductivity tester

Xiangtan Xiangke DRS series high-temperature thermal conductivity tester (guarded heat flow meter method)

I. Overview

This instrument is developed based on our factory's DRS-2 water flow flat plate method thermal conductivity tester. Compared with DRS-2, it improves the resolution and automation, shortens the testing time, is easier to maintain, and has more reliable instrument performance. This instrument is suitable for testing the thermal conductivity of refractory insulation, ceramic fiber, felt, textiles, plates, and bricks at different temperatures. It is widely used in colleges and universities, scientific research units, quality inspection departments, and material analysis and testing in production plants.

This instrument refers to the following standards: ASTM C518-04 Test method for determining steady-state heat flux and heat transfer characteristics using a heat flow meter;  GB10295-2008 Determination of steady-state thermal resistance and related properties of thermal insulation materials (heat flow meter method); YB/T 4130-2005 Refractory materials Thermal conductivity test method (water flow flat plate method); GB/T 17911-2018 Refractory fiber products Thermal conductivity test method, etc.

II. Xiangke DRS-III High-temperature flat plate thermal conductivity meter, thermal conductivity analyzer Main Technical Parameters

DRS series instruments are mainly used for the thermal conductivity detection of thermal insulation materials at high temperatures. DRS-3B has a vacuum atmosphere protection function to prevent sample oxidation.

Instrument Model Specifications

DRS-3

DRS-3A

DRS-3B

1. Thermal conductivity range

0.0010-3W/mk

0.0010-3W/mk

0.0010-3W/mk

2. Test accuracy

±5%

±5%

±5%

3. Hot surface temperature range

100-1200℃

100-1200℃

100-1200℃

4. Calorimetric temperature control method

Constant temperature water bath

Semiconductor refrigeration

Semiconductor refrigeration

5. Heat protection temperature control method

Semiconductor refrigeration

Semiconductor refrigeration

Semiconductor refrigeration

6. Calorimetric plate size

75×75 (mm)

75×75 (mm)

75×75 (mm)

7. Heat flow meter size

75×75 (mm)

75×75 (mm)

75×75 (mm)

8. Number of sample temperature measurement points

4

4

4

9. Sample atmosphere environment

Air

Air

-0.095MPa; Atmosphere protection

10. Sample size

200×200~250×250 (mm); Thickness 10~100 (mm)

11. Measurement and control method

Fully automatic computer testing and data processing, and can generate test reports and print output.

12. Main unit power supply

Voltage 220V; Frequency 50Hz; Total power ≤5KW

13. Instrument configuration

Includes one DRS-3 test host, one software, one set of test thermocouples, one high-precision constant temperature water bath, and one desktop computer (can be provided by the purchaser).

Includes one DRS-3A test host, one software, one set of test thermocouples, one roll of rubber water pipe (the instrument needs to be connected to tap water), and one desktop computer (can be provided by the purchaser).

Includes one DRS-3B test host, one software, one set of test thermocouples, one roll of rubber water pipe (the instrument needs to be connected to tap water), one set of vacuum sealing device, one vacuum pump, and one desktop computer (can be provided by the purchaser).

14. Optional

Multi-layer testing can be selected according to customer requirements. Each additional layer requires an additional test thermocouple, and the cost will be calculated separately.

DRS-4

Both DRS-4A and DRS-4 use electric pressure and automatic thickness measurement. The difference is that the DRS-4 hot plate uses heat-resistant stainless steel, and the operating temperature is slightly lower, while the DRS-4A hot plate uses silicon carbide, the pressure resistance is slightly lower, and the thickness measurement requires an additional thickness correction system, but the hot surface operating temperature is higher.

Instrument Model Specifications

DRS-4

DRS-4A

1. Thermal conductivity range

0.001-3W/mk

0.001-3W/mk

2. Test accuracy

Better than 5%

Better than 5%

3. Hot surface temperature range

200-800℃

200-1200℃

4. Calorimetric temperature control method

Semiconductor refrigeration

Semiconductor refrigeration

5. Heat protection temperature control method

Semiconductor refrigeration

Semiconductor refrigeration

6. Calorimetric plate size

75×75 (mm)

75×75 (mm)

7. Heat flow meter size

75×75 (mm)

75×75 (mm)

8. Number of sample temperature measurement points

4

4

9. Loading and unloading method

Electric

Electric

10. Pressure range

0-1000 (N)

0-500 (N)

11. Thickness measurement range

0-120 (mm)

0-120 (mm)

12. Thickness correction

Experimental correction

Displacement meter instantaneous correction

13. Sample atmosphere environment

Air

Air

14. Sample size

200×200~250×250 (mm); Thickness 10~100 (mm)

15. Measurement and control method

Fully automatic computer testing and data processing, and can generate test reports and print output.

16. Main unit power supply

Voltage 220V; Frequency 50Hz; Total power ≤5KW

17. Instrument configuration

Includes one DRS-4 test host, one software, one set of test thermocouples, one roll of rubber water pipe (the instrument needs to be connected to tap water), and one desktop computer (can be provided by the purchaser).

Includes one DRS-4A test host, one software suite, one set of test thermocouples, one roll of rubber tubing (instrument requires connection to tap water), and one desktop computer (can be provided by the purchaser).

18. Optional Item One

A constant temperature water bath can be optionally selected according to customer requirements, with the circulating water replacing tap water.

19. Optional Item Two

Multi-layer testing can be selected according to customer requirements. Each additional layer requires an additional test thermocouple, and the cost will be calculated separately.

Note: Custom orders for hot surface temperatures of 1500℃ are also available.